Menus, Tabs and Parameters
Settings menu
Wash head
configuration
84
Thermo Scientific Wellwash Versa User Manual
You must always set the wash head configuration to correspond to the
physically installed wash head. Refer to "Head / Plate" on page 63.
Select the Wash head configuration row in the Settings menu and
press the
button.
OK
The Wash head configuration window has the following parameters:
Wash head – Ensure that the installed wash head is the same as set
●
in the Wash head configuration window. Press the
select the correct wash head from the list:
1x8 / 96
•
1x12 /96
•
1x16 / 384
•
2x8 / 96
•
2x12 / 96
•
2x8 / 96 (cell wash)
•
Warning An incorrect wash head configuration can cause liquid spillage
or a faulty washing performance.
Warning If the error message shown below appears, calibrate the wash
head before use.
▲
▲
button to
OK
Thermo Fisher Scientific