Thermo Scientific Wellwash Versa User Manual page 67

Hide thumbs Also See for Wellwash Versa:
Table of Contents

Advertisement

Thermo Fisher Scientific
Note If a 384-well plate with a 1x16 / 384 wash head is selected, you
can select only the Flat well shape.
Note If a 96-well plate with a 2x8 / 96 (cell wash) wash head is selected,
you can select only the Flat, U, and C well shapes.
Note The well shape has an effect on the aspirate height and aspirate
mode.
Flat
C
Flat
Flat bottom
bottom
with curved
edges
Figure 6–36. Well shapes
Well offset – Adjust the X offset of a plate. Adjustment is required
primarily with a 384-well plate. The offset value ranges from
-1.5 mm to 1.5 mm in increments of 0.1 mm.
Caution With 384-well plates, check that the offset limit is not too big
to prevent dispensing outside the wells.
Use the Up and Down arrow keys to adjust the value.
Press the
3 (Preview) key to move the plate carrier and align the wash
F
head over the first column of a plate. Then press the Up or Down
arrow key to move the plate carrier in steps of 0.1 mm.
Warning Well offset has an effect on the residual volume in the well.
With U- and V-shaped wells, setting the well offset off center will
increase the residual volume.
Note With Flat bottom wells, setting the well offset off center may
decrease the residual volume. This feature should only be used with the
Normal aspirate mode.
U
V
V bottom
Round
bottom
Thermo Scientific Wellwash Versa User Manual
Menus, Tabs and Parameters
Main menu
Star A
Star B
Flat bottom
Showing the
with curved
orientation
edges and 8 ribs
of the 8 ribs
65