Menus, Tabs and Parameters
Main menu

Microplate requirements

Well
Default
plate
plate code
96 wells
439454
449824
430341
249662
441653
384 wells
464718
66
Thermo Scientific Wellwash Versa User Manual
Press the
OK
close the window.
Both 96- and 384-well format microplates that meet ANSI/SBS
standards can be used with the Wellwash Versa (Table 6–5).
Note Use only plates manufactured according to ANSI/SBS standard
dimensions.
Table 6–5. Microplate requirements
Default plate
designation
Nunc* solid F96
Nunc U96
Nunc C96
Nunc V96
Nunc starwell strip plate
Nunc F384
* Thermo Scientific™ Nunc™
** Not valid for a 96-well plate with a 2x8 / 96 (cell wash) wash head.

Steps

You can create or edit the wash protocol in the Main menu. It is
recommended to test the optimal parameters for each assay before
operation.
Steps are listed and numbered in the order of execution. It is possible to
add steps to the step list. The maximum number of steps in a protocol is
99.
If there is more than one step in the protocol, the step parameters are
not shown on the Steps row.
Refer to "Adding new steps to protocols" on page 61 and "Deleting
steps from protocols" on page 63.
The following steps are available:
Wash / Cell wash
Aspirate / Cell aspirate
Dispense / Cell dispense
Soak
Prime
Pause
Layout
button to accept the value and/or the
Bottom
Bottom
Dimension name
shape
height
Flat
3.1 mm
Max. overall plate height
U
4.2 mm
C
3.4 mm
V**
4.2 mm
Star**
3.6 mm
Well-to-well distance
Flat
2.7 mm
Max. overall plate height
Well-to-well distance
2 (Close) key to
F
Plate dimensions
14.5 mm ± 0.2 mm
14.5 mm ± 0.1 mm
14.0 mm ± 0.3 mm
14.5 mm ± 0.1 mm
9.0 mm
14.4 mm ± 0.25 mm
4.5 mm
Thermo Fisher Scientific