Maintaining a
system log
How to pack for
Thermo Fisher Scientific
Figure 11–48. Fastening the transport lock
4. Place small plastic bags around the tubes to protect them during
transport.
5. Insert the padded packing material around the plate carrier and
wash head arm for protection after the transport lock is refitted
(Figure 11–49).
Figure 11–49. Transport lock and tag fitted
A system log, which includes a short summary of the use, maintenance
procedures, error messages and other information about the use of the
system, aids proper maintenance of the system. Refer to Appendix A:
"System Log".
To pack the Wellwash Versa for service:
service
Provide information about the use of hazardous materials.
Prime all the channels of the instrument and clean the wash and
waste bottles.
Remove any microplate before decontamination. Decontaminate
the instrument. Remove the liquid bottles and wash heads.

Maintaining a system log

Thermo Scientific Wellwash Versa User Manual
Maintenance