- X-ray tube:
Tungsten X-ray tube
- Exposure X-ray dose: 1 mR
- Maximum size:
IP of 14"x14" (35cm x 35cm) or 14"x17" (35cm x 43cm)
size
- Reference conditions: Distance 1.8 m
Voltage 80 Kv
Amperage 50 mA
Time 0.013 sec
l For "ST-BD" IP
- X-ray tube:
Tungsten X-ray tube
- Exposure X-ray dose: 1 mR
- Maximum size:
IP of 24"x30" ST-BD size
- Reference conditions: Distance 1.8 m
Voltage 80 kV
Amperage 50 mA
Time 0.013 sec
l For "HR" IP
- X-ray tube:
Tungsten X-ray tube
- Exposure X-ray dose: 1 mR
- Maximum size:
IP of 18"x24" HR or 24"x30" HR size
- Reference conditions: Distance 1.8 m
Voltage 80 kV
Amperage 50 mA
Time 0.013 sec
l For "HR-BD" IP
- X-ray tube:
Molybdenum X-ray tube
- Exposure X-ray dose: 20 mR
- Maximum size:
IP of 18"x24" HR-BD or 24"x30" HR-BD size
- Reference conditions: Distance 0.55 m
Voltage 25 Kv
Amperage 100 mA
Time 0.053 sec
011-202-08E
09.30.2009 FM5540
9.2
Image/Conveyance Checks
u NOTE u
Because the "MENU" and so forth differ depending on the IP type, image checks
should be performed with the "MENU" specified below.
1.
Using menus "TEST", "Image Format" or "Image Format-2", and
"SINGLE", record the IP exposed at "9.1 Check Before Procedures",
and generate an image output.
u INSTRUCTION u
For IP type HR-BD, when the image menu is to be registered, the HQ flag should
be set in the "Exposure Parameter Detail Setting".
2.
Check the S value.
REFERENCE
Reference values for S depending on the X-ray dose:
- For IP type ST (tungsten X-ray tube)
For 0.1 mR:
The S value is 2000.
For 1 mR:
The S value is 200.
For 10 mR:
The S value is 20.
- For IP type HR-BD (molybdenum X-ray tube)
For 2 mR:
The S value is 1200.
For 20 mR:
The S value is 120.
For 200 mR:
The S value is 12.