9. Image/Conveyance Checks
In this section, the following checks are made as last-time verification for final
placement.
- Conveyance checks
- Image checks
9.1
Check Before Procedures
u NOTES u
- If a cassette with a wrong IP type is inserted, appropriate image cannot be obtained.
When inserting an IP into a cassette, be sure to perform servicing after confirming
the IP type.
- For the IP used for image/conveyance checks, an IP of the maximum size used in
the hospital should be employed.
- Make sure that the IP to be used does not contain any important image data
exposed.
1.
Check to ensure that the "READY" screen appears.
2.
Perform "Secondary Erasure" on the spontaneous radiation and
image accumulated on the IP to be used.
{Instruction Manual}
011-202-08E
09.30.2009 FM5540
3.
Perform IP exposures.
NOTE
u
u
Be sure to expose the IP under the following X-ray tube voltage condition.
Otherwise, the resulting image could not be checked properly.
For the tungsten X-ray tube:
For the molybdenum X-ray tube: 25 kV
INSTRUCTIONS
u
u
- Expose two IPs of maximum sizes for each of the IP types (ST/HR type, ST-BD
type and HR-BD type) used at the institution, with the following doses.
- With the X-ray tube, the first IP should be exposed in a normal orientation, and
the second in a 180-degree upside-down orientation.
- When the "HR" IP is used in conjunction with the molybdenum X-ray tube, the
IP should be exposed under conditions of "l "For "HR-BD" IP."
CR-IR 363 Service Manual
80 kV
IN-43
IN-43