5
Installation
5.5.4

Connecting the Power Cord

5.5.5

Installing the Nebulizer

Page 62
NOTICE
Condensation in a device can damage the electronics.
Before connecting the devices to the power line, be sure that no
condensation is present in the devices.
If you suspect that condensation is present, allow the device to
warm up to room temperature slowly. Wait until the condensation is
completely gone before proceeding.
1. Verify that the power switch on the device is set to OFF.
2. Connect the power cord to the power inlet connector on the device.
3. Connect the free end of the power cord to an appropriate power
source.
This section describes the installation of the FocusJet nebulizer.
CAUTION—Sharp Tip of the Nebulizer
The capillary inside the nebulizer that protrudes from the nebulizer tip is
sharp and may be fragile. Touching the nebulizer capillary can cause
personal injury and damage to the capillary.
To avoid personal injury, always wear safety glasses when handling
the nebulizer.
Do not touch the nebulizer tip.
Handle the nebulizer with care.
Charged Aerosol Detectors (VH-D20, VF-D20)
Operating Manual