Function Group
Menu Map 8
Description
Programming Codes
See Also
Function Group
Menu Map
Description
Programming Codes
See Also
SYSTEM
This softkey sets the standard starting configuration of the
swept signal generator when the (PRESET) key is pressed, as set
by the user. You can define any starting conditions: Set up the
swept signal generator with the conditions you want, then select
Preset
Mode
User. Now whenever you press[w),the swept
signal generator returns to the configuration you set. If preset mode
user is set, when you press [PJ the swept signal generator
displays the following:
You can still do a factory preset. When the user preset mode
is active, the softkey Factory
[PRESET). An asterisk next to the key label indicates that this feature
is active.
SCPI:SYSTem:PRESet TYPE USER
Analyzer:
NONE
"Changing the Preset Parameters" in Chapter 1
SYSTEM
8
This softkey lets the swept signal generator recognize a printer
address between 0 and 30. The swept signal generator can act as a
controller for a printer during self-test, if the log-to-a-printer feature
is initiated.
SCPI:
DIAGnostic:INSTrument:PRINTer:ADDRess <num>
Analyzer:
NONE
Adrs Menu, Selftest (Full)
Preset appears when you press
Operating and Programming Reference P-l 1
Adrs
Printer