HP 8360 B Series User Manual page 250

Swept signal generator, including options 001, 002, 004, 006, and 008
Table of Contents

Advertisement

Note
Programming Codes
2. If the minimum and maximum frequencies of the swept signal
generator frequency range are not the endpoints desired for the
frequency list, use the Auto Fill Start and Auto Fill Stop
the Auto Fill Incr or Auto Fill #Pts softkeys to create the
list. A list created by this method is ordered with the lowest
frequency as the first point and the highest frequency as the last
point of the frequency list.
Editing Frequency Points
To add a frequency point to the list, place the active entry arrow -->
where you want the next frequency point to appear. The frequency
point is added directly after the value indicated by the arrow.
Lists created by the Auto Fill method are appended to an existing
list much the same way frequency points are added to a list. The
newly created list is added between the frequency point indicated by
the active entry arrow and the point directly after it.
If adding a new list of frequencies causes the existing list to exceed
the maximum number of frequency points allowed (801), the new list
is not appended to the existing list. The error message TOO MANY
LIST PTS REQUESTED is displayed.
To remove a frequency point and its associated offset value and dwell
time, use the delete menu (Delete Current ) softkey. To remove an
entire frequency list, use the delete menu, (Delete All ) softkey.
Editing ALC Offset and Dwell Time
Once a frequency point has been entered, you can assign an ALC
offset and a dwell time value. Use either the Enter List Power
or Global Offset softkey to enter offset values. Use either the
Enter List Dwell or Global Dwell softkey to enter dwell time
values.
The editing softkeys of this menu are not accessible over HP-IB.
Frequency lists to be loaded over HP-IB must first be created in the
controlling program and then downloaded i
swept signal generator.
LIST:FREQuency {<num>[freq suffix]]Iv
LIST:[POWer]:CORRection {<num>[DB]]MAXimum]MINimum}
In the above three commands, the entries contained in {} can be
repeated between 1 to 801 times.
LIST:FREQuency:POINts? [MAXimumIMINimum]
n their entirety to the
Operating and Programming Reference L-7
List Menu