Function Group
Menu Map
Description
Programming Codes
See Also
Function Group
Menu Map
Description
Programming Codes
See Also
FREQUENCY
2
This softkey lets you set the trigger point to be the HP-IB. When
the swept signal generator receives an HP-IB trigger, it steps to the
next frequency point of the frequency list, provided the swept signal
generator is in sweep list mode.
SCPI:
LIST:TRIGger:SOURce BUS
Analyzer:
NONE
List Menu, Pt Trig Menu, Sweep Mode List
"Creating and Using a Frequency List" in Chapter 1
FREQUENCY
2
This softkey lets you set the trigger point to be an external hardware
trigger. When the swept signal generator receives an external
hardware trigger, it steps to the next frequency point of the
frequency list, provided the swept signal generator is in sweep list
mode.
SCPI:
LIST:TRIGger:SOURce EXTernal
Analyzer:
NONE
List Menu, Pt Trig Menu, Sweep Mode List
"Creating and Using a Frequency List" in Chapter 1
Operating and Programming Reference L-9
List Mode Pt TrigExt