Function Group
Menu Map 2
Description
Programming Codes
See Also
F-18 Operating and Programming Reference
FREQUENCY
This softkey lets you set a frequency multiplier value and applies it to
all frequency parameters. Any integer value between and including
it does not affect the output of the swept signal generator.
For example:
1. Set the start frequency to 4 GHz.
2. Set the stop frequency to 10 GHz.
3. Set the frequency multiplier to 5.
Note that the display indicates start=20 GHz, stop=50 GHz and
asterisks appear next to the frequency data.
4. Now set the stop frequency to 30 GHz. The swept signal generator
frequency is 6 GHz, or 30 GHz + 5.
Frequency multiplier and offset are related as shown by the following
equation:
Entered value or Displayed Frequency = (Frequency Generated
The factory preset value is 1. An asterisk next to the key label
indicates that this feature is active.
FREQuency:MULTiplier <num>IMAXimumIMinimum
FREQuency:MULTiplier:STATe ONIOFF(110
<num> will be rounded to the nearest integer.
Analyzer:
SHFA <r-t>
FREQUENCY [MENU], Freq Off set
Multiplier) + Offset value