P R E S E T
Function Group
Menu Map
Description
Programming Codes
See Also
INSTRUMENT STATE
NONE
This hardkey (green) causes the swept signal generator to perform a
short version of self-test, and initializes the swept signal generator to
a standard starting configuration. Two states can be defined for the
standard configuration: Factory or User.
Press (PRESET) at any time to test the swept signal generator and
restore to a standard configuration. If the red LED adjacent to THE
swept signal generator failed self-test; refer to
Swept Signal Generator/HP $360 L-Series Swept CW Generator
Troubleshooting Guide.
Cycling power with the POWER switch does not have the same
effect as presetting the swept signal generator. Cycling power causes
the swept signal generator to display the programming language,
the HP-IB address, and the firmware revision date. After the swept
signal generator displays this data, it restores its configuration to the
state before power was turned off.
SCPI:
SYSTem:PRESet[:EXECute]
Analyzer:
IP
Preset Mode Factory, Preset Mode User
"Changing the Preset Parameters" in Chapter
"Programming Typical Measurements" in Chapter 1
HP 8360 B-Series
1
Operating and Programming Reference P-9