HP 8360 B Series User Manual page 340

Swept signal generator, including options 001, 002, 004, 006, and 008
Table of Contents

Advertisement

l
The specified selftest is executed. Normal instrument operation is
suspended and the instrument state is restored upon exiting the
DIAGnostics:TEST:LOG:SOURce ALLIFAIL
l
DIAGnostics:TEST:LOG:SOURce?
l
Sets and queries the source for the raw data logging. ALL specifies
that all raw data points are displayed. FAIL selects only those data
points out of the test limits. Both commands are executable in
l
l
Selects and queries the raw data logging ON/OFF switch. Both
commands are executable in selftest mode.
After *RST, the setting is 0.
DIAGnostics:TEST:LOOP ONlOFFlllO
l
DIAGnostics:TEST:LOOP?
l
Selects and queries the test looping ON/OFF switch. Both
commands are executable in selftest mode.
After *RST, the setting is 0.
DIAGnostics:TEST:NAME? [<nun01
l
Queries the name of a selftest by number. If the number is not
specified then an array of all the selftest names is returned.
DIAGnostics:TEST:POINts?
l
Returns the number of points of selftest that is output using
DIAGnostics:TEST:NAME?orDIAGnostics:TEST:RESult?.
DIAGnostics:TEST:RESult? [<num>]
l
Queries the result of a selftest, by number. The response is a string
Passed, Failed,
containing either
an array of selftest results are returned.
DIAGnostics:TINT?
l
A test feature that returns the value passed to it. This is used to test
the HP-IB interface.
l
DISPlayC:STATel ONlOFFlllO
l
Sets and queries the display ON/OFF switch.
After "RST, the value is 1.
or
If <num> is missing,
Operating and Programming Reference S-33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83630b83620b83640b83650b83622b83623b ... Show all