4
This softkey (Option 002 only) lets you set the FM deviation for
internally-generated FM. Use the numeric entry keys, arrow keys, or
rotary knob to change the value of the deviation. The swept signal
generator accepts values from 1 Hz to 10 MHz. The factory preset
deviation is 1 MHz.
SCPI:
FM[:DEViation] <num>[freq suffix]]MAXimum]MINimum
Analyzer:
NONE
4
This softkey (Option 002 only) lets you set the FM rate for
internally-generated FM. Use the numeric entry keys, arrow keys,
or rotary knob to change the value of the rate. The swept signal
generator accepts values from 1 Hz to 1 MHz, however it is specified
to 1 MHz only for a sine waveform. Refer to the specifications. The
factory preset rate is 1 MHz (note that the swept signal generator
also presets to a sine waveform).
SCPI:
FM:INTernal:FREQuency <num>[freq suffix]]MAXimum]MINimum
Analyzer:
NONE
(MOD), also see "FM" and "Modulation".