Preset Mode Factory - HP 8360 B Series User Manual

Swept signal generator, including options 001, 002, 004, 006, and 008
Table of Contents

Advertisement

Function Group
Menu Map 8
Description
Programming Codes
See Also
P-10 Operating and Programming Reference
SYSTEM
This softkey sets the standard starting configuration of the swept
signal generator when the (PRESET) key is pressed, as set by the
manufacturer. An asterisk next to the key label indicates that this
feature is active. The following is a description of the configuration.
Start sweep at the minimum specified frequency.
n
Power level set at 0 dBm.
n
Sweep time set to auto.
n
CONT sweep.
n
Sweep mode ramp.
n
ALC leveling point internal.
n
ALC leveling mode normal.
n
Markers set to activate at the center frequency of the sweep.
n
All function values stored in memory registers 1 through 9 remain
n
in their previous states.
The checksum of the calibration data is calculated, and if an error
n
is detected, the calibration data in protected memory is used. If
the checksum of the protected data is not correct, then default
values are used an error message (EEROM FAILED, LOST CAL)
is displayed.
SCPI:
SYSTem:PRESet:TYPE FACTory
Analyzer:
IP, which is the same as (PRESET).
"Changing the Preset Parameters" in Chapter 1