Note
U, V, and W-band power sensors are not available from
Hewlett-Packard. For these frequencies use the Anritsu ML83A
Power Meter with the MP715-004 (40 to 60 GHz), the MP716A
(50 to 75 GHz), or the MP81B (75 to 110 GHz) power sensors. Since
the Anritsu model ML83A Power Meter is not capable of internally
storing power sensor cal factors, you must manually correct the
data entry. Refer to example 2 for information on manual entry of
correction data.
Setup Swept Signal Generator Parameters
5. Turn on the swept signal generator and press (PRESET).
The following occurs:
The source module's frequency span is displayed on the swept
signal generator.
The swept signal generator's leveling mode is automatically
changed from internal to "module leveling".
The source module's maximum specified power is set and
displayed.
6. Press FREQUENCY (START) @ @ 0 @ IGHz), ISTOP) @ @
7. Press
Access User Flatness Correction Menu
8. Press POWER (MENU). Select Fltness Menu.
flatness array is empty.
10. Press [PRIOR). Leave the delete menu and return to the previous
soft key menu.
11. Select Enter Freq @ @ 0 @ (GHz), to enter 26.5 GHz as
the first correction frequency. Enter 31, 32.5, and 40 GHz to
complete the list. Notice that the frequencies are arbitrarily
spaced.
[
]
POWER LEVEL
Getting Started Advanced l-41