Synthesized Step
Sweep
Synthesized List Mode
Ramp Sweep Mode
Internal 10 MHz
Time Base
Accuracy:
Same as time base
Minimum Step Size:
Number of Points: 2
Switching Time:
Same as CW
Dwell Time:
100 /.JS to 3.2 s
Accuracy:
Same as time base
Minimum Step Size:
Number of Points:
1 to 801
Switching Time:
Same as CW
Dwell Time:
100 ps to 3.2 s
Accuracy':
(sweep t ime 2 100 ms and 5 5 s):
Sweep Widths 5 n x 10 MHz: 0.1% of sweep width f time base
accuracy.
Sweep Widths > n x 10 MHz: Lesser of 1% of sweep width or n x
1 MHz + 0.1% of sweep width.
Sweep Time:
10 ms to 100 seconds, 300 MHz/ms maximum rate
Accuracy:
Calibration f Aging Rate & Temperature Effects f Line
Voltage Effects
Stability
Aging Rate: 5 x 10 -"/day, 1 x 10m7/year
With Temperature: 1 x 10-'"/oC, typical
With Line Voltage: 5 x 10 -lo for line voltage change of lo%, typical
Same as frequency resolution
to 801
Same as frequency resolution
Specifications 2c-3