HP 8360 B Series User Manual page 341

Swept signal generator, including options 001, 002, 004, 006, and 008
Table of Contents

Advertisement

S-34 Operating and Programming Reference
FM:COUPling ACIDC
l
FM : COUPling?
l
Sets and queries the FM input coupling mode.
The *RST value is AC.
l
l
Sets and queries the peak FM deviation (in Hz) when the internal
FM generator is used. After *RST, the value is 1 MHz.
FM:FILTer:HPASs <num>[freq suffix'lMAXimumlMINimum
l
FM : FILTer : HPASs? [MAXimum I MINimuml
l
Sets and queries the FM AC bandwidth. There are only two
positions to the bandwidth, < 20 Hz and > 100 kHz, but any
numeric is accepted. The value is compared to 1 kHz and the correct
position is determined (> 1 kHz sets the position to 100 kHz and
100 kHz.
FM:INTernal:FREQuency <num>[freq suffix] 1MAXimumlMINimum
l
FM:INTernal:FREQuency? [MAXimumlMINimuml
l
Sets and queries the frequency (in Hz) of the internal FM source.
After 'RST, the value is 1 MHz.
FM:INTernal:FUNCtion SINusoidlSQUarelTRIaglelRAMPlNOISe
l
FM : INTernal : FUNCt ion?
l
Sets and queries the waveform of the internal FM source.
After *RST, the value is SINusoid.
FM:SOURce INTernallEXTernal
l
FM:SOURce?
l
Sets and queries the source of the FM modulating signal.
After 'RST, the value is EXTernal.
FM:SENSitivity Cnum>Cfreqsuffix/VlIMAXimumlMINimum
l
FM:SENSitivity? [MAXimumlMINimuml
l
Sets and queries the FM Input sensitivity.
The *RST value is MAX (10 MHz/V) .
FM:STATe ONlOFFlllO
l
FM : STATe?
l
Sets and queries the FM modulation state.
After *RST, the value is OFF.