1-42 Getting Started Advanced
Enter Correction Data into Array
is now under swept signal generator control and is performing
the sequence of steps necessary to generate the correction
information at each frequency point.
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the swept signal
generator is using (address 13 is assumed). Refer to the menu
map 8, System, for the key sequence necessary to reach softkey
Meter Adrs .
Enable User Flatness Correction
13. When the operation is complete, (a message is displayed) the
flatness correction array is ready to be applied to your setup.
14. To save the swept signal generator parameters including the
correction table in an internal register, press [SAVE) 0.
(n = number 1 through 8).
15. Disconnect the power meter/sensor and press
(amber LED on). T he p ower produced at the point where the
power meter/sensor was disconnected is now calibrated at the
frequencies and power level specified above.