Sweep Subsystem
Interactions between dwell, sweep time, points, step size, and
frequency span are as follows:
SWEep:TIME = (5 ms + SWEep:DWELl) x (SWEep:POINts - 1)
FREQ:SPAN = SWEep:STEP x (SWEep:POINts - 1)
T I M E
OFF
OFF
OFF
ON
ON
OFF
ON
ON
SWEep:CONTrol:STATe ONIOFFIllO
l
SWEep:CONTrol:STATe?
l
Sets and queries the state of the sweep control.
Normal source mode.
OFF
Use master slave source mode.
O N
*RST value is OFF.
SWEep:CONTrol:TYPE MASTerlSLAVe
l
Sets and queries the swept signal generator, whether it is in master
or slave mode. This applies in a dual source mode. *RST value is
l
l
Sets and queries the amount of time in seconds that the swept signal
generator stays (dwell) t a each step after reporting a
SRQ
and pulsing the Trigger
each step when in the SWE:TRIG: SOUR IMM mode of a stepped sweep.
Setting SWEep:DWELL sets SWEep:DWELl :AUTO OFF. *RST value is
100 /Js.
l
l
Sets and queries the state of the automatic dwell calculation switch.
Setting SWEep:DWELL sets SWEep:DWELl:AUTO OFF. "RST state is OFF.
S W E e p : x x : A U T O s w i t c h e s
Switch Action
No coupling between SWEep:DWELl,
and SWEep:POINts.
No coupling between SWEep:DWELl,
SWEep:TIME and SWEep:POINts.
When SWEEP:TIME or SWEEP:POINts
are changed, SWEep:DWELl =
(SWEep:TIME / (SWEep:POINts - 1)) - 5 mS.
SWEep:DWELl is limited to 100 ps, minimum.
SWEep:DWELl = 100 ps (MINimum).
SWEep:TIME = 5.1 ms
Out
line low. This one value is used at
Operating and Programming Reference S-49
(SWEep:POINts - 1)
x
source settled