HP 8360 B Series User Manual page 351

Swept signal generator, including options 001, 002, 004, 006, and 008
Table of Contents

Advertisement

S-44 Operating and Programming Reference
ON
ON
ON
OFF
ON
ONCE
not applicable
OFF
POWer : SLOPe <num> [DB/f req
l
POWer : SLOPe?
l
Sets and queries the RF slope setting (dB per Hz).
FREQ:MODE Affect on Slope
CW or LIST
Rotates around 0 Hz.
Rotates around the start frequency.
STEP
The *RST value is 0.
POWer:SLOPe:STATe ONlOFFlllO
l
POWer:SLOPe:STATe?
l
Sets and queries the power slope state. *RST value is 0.
POWer:SPAN <num>[DB] ~MAXimumlMINimumlUPlDOWN
l
POWer : SPAN? [MAXimum I MINimum]
l
The coupling equations for power sweep are exactly analogous to
those for frequency sweep. Power sweep is allowed to be negative,
unlike frequency sweeps. *RST value is 0.
POWer:STARt <num>[lvlsuffix] lMAXimumlMINimumlUPlDOWN
l
POWer:STARt? [MAXimumlMINimum]
l
Default units and units for query response are determined by the
command UNIT:POWer. The coupling equations for power sweep
are exactly analogous to those for frequency sweep. Power sweep is
allowed to be negative, unlike frequency sweeps.
*RST value is 0 dBm.
POWer:STATe ONlOFFlllO
l
POWer : STATe?
l
Sets and queries the output power on/off state. *RST value is OFF.
P O W e r : S E A R c h
ALC is momentarily closed to
level at the requested power,
and then the modulator is set to
the same voltage in open loop
mode. This repeats automatically
any time that the power level or
frequency is changed.
Normal mode.
Immediately performs a power
search. This leaves POWer:SEARch
in the ON position.
Modulator setting is
explicitly set by user.
Power Switch Action
I