Function Group
Menu Map
Description
Programming Codes
See Also
Function Group
Menu Map
Description
Programming Codes
See Also
S-70 Operating and Programming Reference
FREQUENCY
2
When this softkey is selected, the swept signal generator
automatically steps to the next point in the stepped frequency sweep
until all points are swept. The time between points is equal to the
sum of the dwell and phase lock times. An asterisk next to the key
label indicates that this feature is active.
SCPI:
SWEep:TRIGger:SOURce IMMediate
Analyzer:
NONE
Step Sup Menu, Sweep Mode Step
"Using Step Sweep" in Chapter 1
FREQUENCY
2
When this softkey is selected, the swept signal generator steps to the
next point in a stepped frequency sweep when an HP-IB trigger
(*TRG, <GET>) is received (leading edge TTL). When the last
frequency point is reached and continuous sweep is selected, the
next trigger causes the step sweep to return to the start frequency.
Connect the trigger signal to the TRIGGER INPUT BNC. An
asterisk next to the key label indicates this feature is active.
SCPI:
SWEep:TRIGger:SOURce BUS
Analyzer:
TS
Step Sup
Menu , Sweep Mode Step