Causes the sweep in progress to abort and
ON it immediately restarts the sweep. The pending operation flag
(driving *OPC, *WA& and *OPC?) undergoes a transition once the
sweep is reset.
l
l
AM[:DEPThl? [MAXimumIMINimuml
Sets and queries the percentage of AM when the SOURce is
converted to percent by the equation:
Valid ranges of dB are 0 to 40 dB. After *RST, the value is 50%.
l
l
AM:INTernal:FREQuency? [MAXimumIMINimum]
Sets and queries the frequency (in Hz) of the internal AM source.
After *RST, the value is 1 kHz.
AM:INTernal:FUNCtion SINusoidlSQUarelTRIanglelRAMPlNOISe
l
l
AM:INTernal:FUNCtion?
Sets and queries the waveform of the internal AM source. After
l
l
AM:SOURce?
Sets and queries the source of the AM modulating signal. After
AM:MODE DEEPlNORMal
l
Controls the AM depth limits of the swept signal generator. The
AM:STATe ONlOFFlllO
l
AM:STATe?
l
Sets and queries the status of the AM modulation. After *RST, the
setting is OFF.
AM:TYPE LINearlEXPonential
l
AM:TYPE?
l
Sets and queries the type of AM modulation. After *RST, the
setting is LINear.
CALibration:AM:AUTO ONIOFFIlIO
l
CALibration:AM:AUTO?
l
Sets and queries the automatic modulator calibration switch.
reset. If INIT: CONT