Controlling the DI probe temperature

Source temperature is often sufficient to evaporate the sample, particularly if
the shallow sample cup is used. If the sample fails to evaporate, or if a more
controlled evaporation is required, MassLynx can ramp the probe
temperature during acquisition.

Ramping the probe temperature manually

To ramp the probe temperature manually:
1.
In the MassLynx Tune window Inlet tab, ensure that the Use DCI mode
check box is not selected.
Tune window Inlet tab:
2.
Enter the required temperature in the Solids Probe text box.
3.
Press the Enter key.
Alternative:
Tip:
For safety reasons, the temperature is set to 0 at the start of an
acquisition. Once the acquisition has started, you can manually ramp
the probe temperature using this control.
7-10
The Direct Insertion Probe
Use the slider control to select the required temperature.