Monitoring The Instrument Vacuum - Waters GCT Premier Operator's Manual

Table of Contents

Advertisement

To vent the source:
1.
In the MassLynx Tune window, click Vacuum > Vent Source.
2.
Wait for the Vacuum Status LED to display steady yellow.
See also:
Result:

Monitoring the instrument vacuum

The instrument's vacuum status appears in the Vacuum Monitor window,
in mbar. Readbacks from the instrument's pirani gauge on the backing pump
and the penning gauges in the analyzer and the source are given.
If any Pirani gauge output pressure reading exceeds the trip level, the
relevant fine pump is turned off. There is some hysteresis in the operation of
this trip, thus the pressure at which the trip resets when the vacuum is
improving is lower than that at which it trips out.
The Engineer tab on page
See also:
To monitor the instrument's vacuum levels:
In the MassLynx Tune window, click
window.
Alternative:
Note:
The following illustration shows typical values (in mbar) for an
instrument when fully pumped-down with a GC fitted and helium flowing at
1 mL/min.
Vacuum monitor window:
Status display on page
The source is now vented.
Click Vacuum > Monitor.
1-9.
B-39.
to open the Vacuum Monitor
The Tune window Vacuum menu
B-9