Probe Control dialog box Options tab parameters:
Parameter
Use Ramps
Use Base Peak
Use TIC control
On Threshold
Hold Threshold
Abort Threshold
After Mass
Ramping Property
Name pane
7-12
The Direct Insertion Probe
Description
Enables probe ramp control.
Enables probe temperature regulation based on the
base peak intensity. The base peak is the most
intense peak within the scanned mass range.
Enables probe temperature regulation based on the
Total Ion Current (TIC).
When temperature regulation is enabled, this param-
eter specifies the intensity (of the current property
used for regulation: for example, Base Peak or TIC).
At that, the probe ramp intensifies will increase the
probe temperature after a hold or abort event.
Tip:
The specifies value is in counts.
Specifies the intensity at which the probe ramp
control will hold the probe temperature until the
intensity of the current property reaches the On
Threshold or Abort Threshold.
Specifies the intensity at which the probe ramp
control is turned off until the intensity reaches the On
Threshold.
If Use Base Peak is selected, masses below the value
entered in the associated Mass text box are ignored
when calculating the base peak value.
Displays which device will be ramped, and which of
its properties will be controlled.
Depending on which probe is selected, the
Tip:
parameters displayed.
See also:
Figure titled "Tune window Inlet tab:" on
page
7-10.