Detector Maintenance; Conditioning The Detector - Waters GCT Premier Operator's Manual

Table of Contents

Advertisement

4.
Connect the CI gas supply to the CI gas connector on the rear of the
instrument.
See also:
5.
Allow the backing pirani read back to reach 1.00e-2 mbar. If this cannot
be achieved note the read back. (Sometimes a minor amount of
condensation within the line prevents attaining the required pressure.)
6.
In the Tune window Inlet tab, click CI purge.
Result:
should recover back to the initial backing pressure within 30 seconds.
7.
Wait for approximately 10 minutes.
8.
Click CI Purge.
Result:
does not recover to the initial backing pressure in
pressure kick after the second purge in
must be located and corrected.
9.
Repeat

Detector maintenance

Conditioning the detector

The detector must be conditioned before use. You condition the detector by
gradually increasing an applied voltage over a period of time.
Caution:
vented. Failure to follow the detector conditioning procedure will
reduce lifetime of the detector and may destroy it.
To condition the MCP detector:
1.
Ensure the analyzer and source are pumped down (evacuated). The
analyzer vacuum must be <3e-6 mbar for one hour before the
conditioning starts.
2.
In the MassLynx Tune window Source tab, ensure that the Detector
Voltage is set to 0.
10-50 Routine Maintenance
Gas inlet connections on page
The read-back should increase to 1.00e0 mbar. The pressure
There should not be any kick in the pressure. If the read-back
step 5 through step
The detector must be conditioned each time the analyzer is
1-19.
step
8, a leak is present and
8, until no leaks are present.
step 5
or if there is a