Changing the EI/CI outer source assembly

Removing the EI/CI outer source from the instrument

To remove the EI/CI outer source:
1.
On the MassLynx Tune Window, click Vacuum > Vent Source.
2.
Wait for the Vacuum Status LED to display steady yellow.
Warning:
interface to cool before handling.
3.
If fitted, remove the GC column from the instrument.
See also:
4.
Move the GC away from the right side of the instrument.
5.
Remove the inner source from the outer source.
See also:
the FI/FD inner source on page
2-6
Preparing the instrument for EI/CI operation
The source component scan be contaminated with
Warning:
toxic and biohazardous materials. Wear chemical-resistant,
powder-free gloves at all times while handling the
components.
The source is hot and can cause burn injuries to
Warning:
the skin. Allow the source to cool before handling it.
To avoid burn injuries, allow the GC column and
Installing the GC interface on page
Removing the EI/CI inner source on page 2-3
6-3.
9-13.
and
Removing