2.
Sonicate the lens plates in a suitable solvent to remove any remaining
grease.
See also:
Ultrasonically cleaning the inner source components on
page
10-21.
Removing the collimating slits
The two collimating slits are 0.25 mm wide and can be removed from the outer
source when using FI or FD ion mode.
Note:
When the collimating slits are removed from the outer source there is a
noticeable improvement in sensitivity, but reduction in resolution.
To remove the collimating slits:
1.
Unscrew and remove the securing screws that secures the slit assembly
onto the slit mounting block.
2.
Lift up and remove the slit assembly from mounting block.
Collimating slits:
3.
Rub each slit, with 1200-grade wet/dry paper to remove any visible ion
burn.
Securing screw
Removable slit
assembly
Slit mounting block