Starting-up the instrument

To start-up the instrument:
1.
Ensure the power switch is switched on at the main power supply.
2.
Ensure the compressed air is set to 100 psi.
See also:
3.
Ensure the CI gas is set to 10 psi.
4.
Switch on the instrument power on/off switch at the rear of the
instrument.
5.
Switch on the MassLynx PC.
6.
Switch-on the embedded PC (EPC).
See also:
7.
Allow 3 minutes for the embedded PC to initialize.
8.
Start the MassLynx software.
9.
In the MassLynx window, click
See also:
10. Switch on the backing pump.
See also:
11. In the Tune window, click Vacuum > Pump Instrument.
Result:
select Yes,
Waters recommends that you condition the detector for 10 hours
Tip:
when the analyzer has been vented to air. If the analyzer has been
vented to dry gas and not exposed to air, conditioning can be reduced to
60 minutes.
See also:
12. Click the Engineer tab.
13. Ensure that the instrument has pumped sufficiently so that the Vacuum
LED on the front panel is green.
See also:
A-2
Starting-up and Shutting-down the Instrument
Gas inlet connections on page
EPC power switch on page
Opening the MassLynx Tune window on page
Rotary pump maintenance on page
You are prompted to auto-condition the MCP detector. If you
steps 13 to 15
are completed automatically.
Conditioning the detector on page