Changing the EI/CI inner source assembly

Caution:
column and inner source chamber assembly, the source components
must be removed in the order shown in the
the source components:" on page
components, they must be installed in the reverse order.
Removing the source components:
Inner source
EI, CI, DCI, FI, FD
Order of fitting or removing source components:
Fitting
1
Outer source
2
Inner source
3
GC or, DI probe or, DCI probe Outer source
2-2

Preparing the instrument for EI/CI operation

If the GC column is installed, to avoid damaging the GC
Outer source
EI/CI/DCI, FI/FD
Figure titled "Removing
2-2. When installing the source
GC column + interface
or
DI probe
or
DCI probe
Removal
GC or, DI probe or, DCI probe