Cleaning The Complete Outer Source - Waters GCT Premier Operator's Manual

Table of Contents

Advertisement

2.
Ultrasonically clean the slits in a suitable solvent to remove any
remaining grease.
See also:
page
10-21.
To refit the collimating slits:
1.
Assemble the collimating slits assembly in the reverse order to which
you disassembled it.
2.
Fit the collimating slits assembly into the outer source assembly.
3.
Fit and tighten the screw that secures the collimating slits to the outer
source lid assembly.
4.
Fit the outer source lid assembly to the instrument.
See also:

Cleaning the complete outer source

To disassemble the outer source:
1.
Remove the outer source lid assembly from the instrument.
See also:
page
2-6.
2.
Place the outer source lid assembly on a bench so that the front support
plate is at the top.
3.
Make a diagram of where the wires, thermocouple, and gas tubing are
located.
Rationale:
4.
Unscrew and remove the two hex-head screws that secure the front
support plate to the outer source lid.
Ultrasonically cleaning the inner source components on
Fitting the EI/CI outer source to the instrument on page
Removing the EI/CI outer source from the instrument on
This diagram aids assembling the outer source.
EI/CI outer source maintenance 10-37