Waters GCT Premier Operator's Manual page 178

Table of Contents

Advertisement

FI/FD inner source lock assembly:
6.
Ensure that the source isolation valve is closed.
When the outer source and the inner source lock assembly has been
fitted you can pump the instrument down.
7.
Open the reference reservoir valve.
8.
In the MassLynx Tune window, click Vacuum > Pump Source.
When the pneumatic valve opens, you should hear a small hiss.
9.
Ensure that the Vacuum LED shows flashing green followed by steady
green.
See also:
10. When the Source Turbo value in the Turbo Readbacks frame reaches 80
or more, close the reference reservoir valve.
9-8
FI/FD Operation
Status display on page 1-9.
Source isolation valve
(closed position)
Hex-head retaining
screw (captive)
Vent Valve

Advertisement

Table of Contents
loading

Table of Contents