Physical Specifications
Table 3
Type
Weight
Dimensions
(height × width × depth)
Line voltage
Line frequency
Power consumption
Ambient operating
temperature
Ambient non-operating
temperature
Humidity
Operating altitude
Non-operating altitude
Safety standards:
IEC, CSA, UL
Hot rear panel
WA R N I N G
Using the module at high environmental temperatures may cause the rear panel to
become hot.
➔
Agilent 1260 Infinity Standard Autosampler User Manual
Physical Specifications
Do not use the module at environmental temperatures higher than 50 °C (122 °F)
Site Requirements and Specifications
Specification
14.2 kg (32 lbs)
200 × 345 × 435 mm
(8 × 13.5 × 17 inches)
100 – 240 VAC, ± 10 %
50 or 60 Hz, ± 5 %
300 VA / 200 W / 683 BTU
0–55 °C (32–131 °F)
-40 – 70 °C (-40 – 158 °F)
< 95 % r.h. at 40 °C (104 °F)
Up to 2000 m (6562 ft)
Up to 4600 m (15091 ft)
Installation category II, Pollution degree 2
Physical Specifications
Comments
Wide-ranging
capability
Maximum
See warning
"Hot rear
panel"
on page 25
Non-condensing
For storing the module
For indoor use only.
2
25