Physical Specifications - Agilent Technologies InfinityLab LC Series User Manual

Preparative open-bed fraction collector
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

2
Site Requirements and Specifications

Physical Specifications

Physical Specifications

Table 1
Type
Weight
Dimensions
(height × width × depth)
Line voltage
Line frequency
Power consumption
Ambient operating tem-
perature
Ambient non-operating
temperature
Humidity
Operating altitude
Safety standards: IEC,
EN, CSA, UL
ISM Classification
Permitted solvents
1290 Infinity II Preparative Open-Bed Fraction Collector User Manual
Physical Specifications
Specification
30.6 kg (67.4 lbs)
781 mm x 393 mm x 622 mm
(30.7 inches, 15.5 inches, 24.5 inches)
100 – 240 V~, ± 10 %
50 or 60 Hz, ± 5 %
350 VA, 350 W
4-40°C (39 -104°F)
-40 – 70 °C (-40 – 158 °F)
≤80 % r.h. up to 31 °C, decreasing to
50 % r.h. at 40 °C
Up to 3000 m (9842 ft)
Installation category II, Pollution
degree 2
ISM Group 1 Class B
Boiling point ≥56 °C (133 °F).
Auto-ignition temperature ≥200 °C
(394 °F).
Comments
Wide-ranging capability
Non-condensing
For indoor use only.
According to CISPR 11
Do not use any solvent that is
not explicitly mentioned on the
"solvent compatibility" list.