Command Summary - Keithley 2182 Service Manual

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

Command summary

Table B-1
detail in the following paragraphs.
Table B-1
Remote calibration command summary
Command
:CALibration
:PROTected
:CODE <name>
:INITiate
:PCAL
:STEP0 <nrf>
:STEP1 <nrf>
:STEP2 <nrf>
:DC
:STEP1
:STEP2
:STEP3 <nrf>
:STEP4 <nrf>
:STEP5
:DATE <string>
:DATE?
:NDUE <string>
:NDUE?
:SAVE
:LOCK
:LOCK?
:DATA?
:COUNt?
**Factory cal only. Send :DIAG:KEIT:CAL:UNLOCK to enable.
**Calibration data will not be saved if an error occurred.
summarizes Model 2182 calibration commands. These commands are covered in
Description
Send the code to unlock calibration. Default: 'KI002182'.
Required before starting any other cal steps.
Precal commands.*
Precal 10V input.
Precal 0V input.
Precal 1V front input.
Normal calibration commands.
Normal calibration short-circuit input.
Normal calibration open-circuit.
Normal calibration +10V input.
Normal calibration –10V input.
Analog output calibration.*
Set the cal date.
Query cal date.
Set the due date for the next cal.
Query cal due date.
Save cal constants.**
Lock out calibration.
Query lock status.
Query cal constants (unlocking not required).
Query cal count.
Calibration Reference
B-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

2182a

Table of Contents