To Change The Glass Liner On The Purged Packed Inlet - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

6
Maintaining the Purged Packed Inlet

To Change the Glass Liner on the Purged Packed Inlet

WA R N I N G
Top insert weldment
Flared end
Glass liner
C A U T I O N
84
1
Gather the following:
Replacement glass liner (See
for the Purged Packed Inlet"
9/16- inch wrench
Lint- free gloves
Load the
inlet maintenance method
2
to become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Loosen the top insert weldment to remove the top portion
3
of the inlet.
4
Use a thin wire or wood splint to carefully lift and
remove the old glass liner.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
"Consumables and Parts
on page 68.)
and wait for the GC