Temperature Input Unit Functions
Checking the Wiring and
Making Adjustments
1,2,3...
Note
Input Terminal Connections
Wiring for Adjusting the Upper and Lower Limits
DMM
Ferrule for two wires
NC
0
NC
1
+
+
NC
0
NC
1
Cold junction compensator
STV
Lead wires for adjusting bias
NC
0
NC
1
+
+
NC
0
NC
1
Cold junction compensator
Compensating conductors
INPUT
Adjusting the Upper and Lower Limit Values
1. Set the wiring for adjusting the upper and lower limits. If connecting the
precision digital multimeter (DMM), use a ferrule for two wires.
2. Check the sensor and input type being used.
When using an R, S, B, E, or W sensor, use a compensating conductor for a K
thermocouple. In addition, when using an R, S, or B sensor, set the input type
to K (0.0 to 500.0 ° C). When using an E or W sensor, set the input type to K
( − 200 to 1,300 ° C).
3. Connect the CX-Integrator to the CompoNet network and go online.
4. Upload settings to the CX-Integrator.
5. Turn ON the power supplies of all Units, including the Temperature Input
Unit to be adjusted. Wait approximately 30 minutes for the Temperature In-
put Unit's internal temperature to stabilize.
6. Double-click the icon of the Temperature Input Unit to be set in the Network
Configuration Window or right-click the icon and select Parameters - Edit.
The Configuration Window will be displayed.
7. Click the tab for the input that will be adjusted and click the Adjustment
Button.
NC
NC
2
NC
+
2
NC
3
NC
NC
2
NC
3
+
+
2
NC
3
ZERO-CON (0°C bath)
Short-circuit
0°C/32°F
OUTPUT
Section 11-5
3
+
481