Analog Output Terminal Block - Omron ZW-7000 Series User Manual

Confocal fiber type displacement sensor
Hide thumbs Also See for ZW-7000 Series:
Table of Contents

Advertisement

Analog output terminal block

Used for analog output.
Compatible cable specifications: Solid wire 0.2 to 1.5 mm
Class
Name
Analog
OUT (V)
output
OUT (A)
OUT 0V
Important
• Cut the unnecessary signal lines so that they do not touch other signal lines.
Make the wire length as short as possible. Do not use a wire length of over 30 m.
• Analog output is not possible under the following combination of conditions.
Measurement cycle is less than 40 µs.
EtherCAT output is enabled.
• Make sure to fix the terminal block to prevent be injure when pushing the release button with a screw driver.
72
Wiring
2
Stranded wire 0.2 to 1.5 mm
Stranded wire with rod terminal(no plastic sleeve) 0.25 to 1.5 mm
Stranded wire with rod terminal(with plastic sleeve) 0.25 to 0.75 mm
AWG 24 to 16
Pin processed length 10 (±0.5) mm
Signal name
No.
OUT(V)
1
OUT(A)
2
OUT 0V
3
Description
This outputs the measured value, from -10 V to +10 V as the voltage value.
When measurement not possible: Approx. +10.8 V (default value; can be selected by user)
Alarm: Approx. 10.8 V
This outputs the measured value, from 4 mA to 20 mA as the current value.
When measurement not possible: Approx. +20.8mA (default value; can be selected by
user)
Alarm: Approx. +20.8 mA
0 V for analog voltage output.