Executing Sample Programs
HCI Measurement Program

Setting up the Measurement Environment

Before executing the program, you need to set up the measurement environment as
follows:
1. Prepare the equipment shown in Table 9-6, which lists the required equipment to
execute the program.
2. For each E5255A, make internal connections, and set the DIP switches. Refer to
Figure 9-3.
Three E5255As are used, which are installed in Slot1, Slot2, and Slot3 of the
E5250A. The internal connections of the three E5255As are almost the same,
except for the following:
IVin1 internal connection (1 in Figure 9-3) for each E5255A:
for Slot1: IVin1 to SMU1
for Slot2: IVin1 to SMU3
for Slot3: IVin1 to SMU4
DIP SW settings in each slot are different as shown in Figure 9-3. But within
each E5255A, SW1/SW2/SW3 settings are the same.
3. Connect all equipment and devices as shown in Figure 9-4 and Table 9-7. Refer
to Chapter 4 for the details of the connection from the E5250A output to Device
Under Test (DUT).
9-24
Agilent E5250A User's Guide, Edition 11