:DIAGnostic:TEST:FRAMe[:EXECute]?
:DIAGnostic:TEST:FRAMe[:EXECute]?
This command executes the specified E5250A test, then returns the test result. For
more info, see Chapter 3.
After :DIAG:TEST:FRAM:EXEC CONT execution, the E5250A status becomes
same as after *RST command execution.
After :DIAG:TEST:FRAM:EXEC FPAN execution, the E5250A status is not
changed.
The :DIAG:TEST:FRAM:EXEC FPAN command requires key input from the
front panel. If key is not pressed within 10 seconds, this test result will be fail.
:DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel
Parameter
CONTroller
FPANel
test_result <newline><^END>
1 : Fail
0 : Pass
OUTPUT @Hp5250;":DIAG:TEST:FRAM? CONT"
7-18
Explanation
Controller test
Front panel interface test