Error Messages
This chapter lists and describes the error messages for Agilent E5250A. An error
message consists of an error number and message.
There are two types of error messages:
Standard SCPI Error Messages
Negative error numbers (Command Error, Execution Error, Device-Dependent
Error, and Query Error) are standard SCPI errors.
E5250A Specific Error Messages
Positive error numbers are the E5250A specific errors.
Error messages are classified by error number as listed in the following table:
Error Range
0
-100 to -199
-200 to -299
-300 to -399
-400 to -499
100 to 32767
When an error occurs, the corresponding bit is set in the Standard Event Status
Register (see Chapter 7):
Error Category
Command Error
Execution Error
Device-Dependent Error
Query Error
E5250A Specific Error
Also, the error number and message are placed in the error queue, which can be read
by the :SYSTem:ERRor? query command. The error queue is cleared by the
common command *CLS, and when power is turned on. For these commands, see
Chapter 7.
11-2
Error Category
No error
Command Error
Execution Error
Device-Dependent Error
Query Error
E5250A Specific Error
Standard Event
Status Register Bit
Agilent E5250A User's Guide, Edition 11
bit5
bit4
bit3
bit2
bit3