Example:
Programming the E5250A
Creating a Control Program

Controlling Relay Switching

Following commands are used to control relay switching. For details of
channel_list, see "Channel List and Channel Configuration" on page 5-6.
Command
:ROUT:CLOS channel_list
:ROUT:OPEN channel_list
:ROUT:OPEN:CARD card_no
100 OUTPUT @Hp5250;":ROUT:CLOS (@10101,10202)"
110 ! Closes relays to create channels 10101 and 10202.
6-12
Description
Close the relays specified by channel_list.
Open the relays specified by channel_list.
Open all relays on card specified by card_no.
Agilent E5250A User's Guide, Edition 11