Agilent Technologies Agilent E5250A User Manual page 173

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

Syntax
Example
Syntax
Query response
Semantics
*CLS
This command clears the Status Byte Register, the Standard Event Status Register,
and the Error Queue. This command does not clear the enable registers. Refer to
"Status Reporting Structure" on page 7-48.
Also, this command stops the monitoring of pending operations by the *OPC
command.
This command does not have query form.
*CLS
OUTPUT @Hp5250;"*CLS"
*ESE
This command sets the bits of the Standard Event Status "Enable" Register.
1 enables, 0 masks.
*ESE enable_number
Parameter
enable_number
decimal integer (that is the sum of the binary-weighted values
for the desired bits), hexadecimal, octal, or binary value
enable_number <newline><^END>
enable_number is decimal integer value that is the sum of the binary-weighted
values of the "Enable" register bits.
The Standard Event Status "Enable" Register determines which bits of the Standard
Event Status Register are enabled. Enabled bits are ORed together, and the result is
reported to bit5 of the Status Byte Register.
The Standard Event Status "Enable" Register consists of 16 bits, but only the lower
8 bits are used, which correspond to the bits of the Standard Event Status Register.
For details, refer to "Standard Event Status Enable Register" on page 7-54
Agilent E5250A User's Guide, Edition 11
Command Reference
Explanation
*CLS
7-5