Figure 6-3
Required
Conditions:
Couple Port Example
The following example shows how to use the Couple Port.
This function is available only for the E5252A.
Example of Couple Port
This program assumes the following conditions.
E5250A Installed Cards:
•
Slot 1: E5252A must be installed.
•
Slot 2 to 4: not used.
E5250A Input Ports:
•
SMU INPUT 1 and 2: should be connected to instrument.
•
SMU INPUT 3 &4 and 5&6: should be connected to instrument. Couple port.
•
AUX INPUTs: not used.
The program couples SMU INPUT 3&4 and couples SMU INPUT 5&6. For each
Couple Port, use Kelvin triaxial cable for instrument connection.
E5252A Output Ports:
Output 1, 2, 3, 4, 5, 6 should be connected to :
Substrate, Gate, Drain, Drain, Source, Source, respectively.
Agilent E5250A User's Guide, Edition 11
Programming the E5250A
Programming Examples
6-19