Agilent Technologies Agilent E5250A User Manual page 82

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

NOTE
Setting up Measurement Environment
Mounting Connectors Directly
To Check Interlock Circuit
If you use the 4155/4156, you can easily check the interlock circuit as follows:
1. Connect the Intlk connector of the 4155/4156 to your interlock circuit.
2. Press
front-panel key, then select CALIB/DIAG primary softkey to
System
display the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS page.
3. In the CALIB/DIAG field, select DIAG secondary softkey.
4. In the CATEGORY field, select I/O PERIPH secondary softkey.
5. Move pointer to the 403 (INT.) Interlock LED field.
6. Select EXECUTE secondary softkey.
7. Confirm the following:
LED turns on within 1 sec from when interlock circuit is shorted.
LED turns off within 1 sec from when interlock circuit is open.
To stop the interlock test, select STOP secondary softkey.
4-12
Agilent E5250A User's Guide, Edition 11