Agilent Technologies Agilent E5250A User Manual page 301

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

Typical Data
*1
Offset Current
Channel Crosstalk Capacitance < 0.4 pF per channel (Low Leakage I-V Port)
Offset Voltage, EMF at 5min
Guard Capacitance
*1
The offset current when 0 V is applied to all I/O channels.
*2
EMF means Electro Motive Force, and these are offset voltages after
applying voltage and waiting 5 minutes.
*3
This is guard capacitance for an I/O channel for following conditions.
Only one E5255A is installed in mainframe, and only one I/O channel
is connected.
Supplemental Data
Relay Contact Life:
Settling Time:
Agilent E5250A User's Guide, Edition 11
< 0.1 pA (Low Leakage I-V Port)
< 5 pF per channel (Bias Port)
*2
< 80 μV (Low Leakage I-V Port)
*3
< 180 pF
(Low Leakage I-V Port)
< 60 pF (Low Leakage I-V Port, when the port
is open.)
8
> 10
times (Dry Switching Mode)
< 3.5 sec to settle to within 400 fA of final value
after 10 V is applied (Low Leakage I-V Port)
Specifications