Executing Sample Programs
HCI Measurement Program
To change the 4155/4156 setup file used
To change the 4155/4156's setup file used for the HCI test program, modify the
setting of the setup file and save it as a new setup file. Then change the following
parameter in the "Definition of measurement and stress setup files" block of the
program.
Parameter
Ibvg_file$
Setup file for extracting Vgstr.
Igleak_file$
Setup file for Gate leakage current
measurement.
Idleak_file$
Setup file for Drain leakage current
measurement.
Param_file$
Setup file for parameter measurement.
To change Vtci extraction condition
To change W/L ratio (Gate width/Gate length ratio) required for Vtci extraction, do
follows.
1. GET the PARAM.MES setup file by using the 4155/4156.
2. Change the setting of "ID" parameter in "LINE2" on the ANALYSIS SETUP
screen. The value of "ID" can be calculated by the following expression.
ID=1 μA × W/L
where, W means Gate width in m, and L means Gate length in m.
3. Change the following parameters in the "Device geometries" block of program.
Parameter
Gate_length
Gate_width
9-36
Description
Description
Gate length in m.
Gate width in m.
Agilent E5250A User's Guide, Edition 11
Default
"IBVG.MES"
"IGLEAK.MES"
"IDLEAK.MES"
"PARAM.MES"
Default
1.E-6
1.E-5