Programming the E5250A
Programming Examples
160 !
170 ! Put Bias Source Control Routine here.
180 !
190 OUTPUT @Hp5250;":ROUT:CLOS "&Channel$
200 OUTPUT @Hp5250;"*OPC?"
210 ENTER @Hp5250;Complete
220 !
230 ! Put Measurement Routine here
240 !
250 OUTPUT @Hp5250;":ROUT:BIAS:STAT ALL,OFF"
260 OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"
270 OUTPUT @Hp5250;"*OPC?"
280 ENTER @Hp5250;Complete
290 !
300 OUTPUT @Hp5250;"*RST"
310 !
320 END
Line
Number
170
Here, insert your bias source control routine.
190
Disconnects all output ports of block 1 from the input Bias Port,
then connects the output ports to the IV input port.
200 to 210
Waits until the channel connections are completed.
230
Here, insert measurement routine for your measurement
instrument.
250
Turns off the Bias Mode.
260 to 280
Opens all relays, then connects all output ports of block 1 to the
input Bias Port again, and waits until the connections are
completed.
300
Resets the E5250A.
6-18