Agilent Technologies Agilent E5250A User Manual page 133

Agilent technologies low leakage switch mainframe user's guide
Table of Contents

Advertisement

NOTE
To enter a comment for the setup memory
Select the COMMENT softkey. The following message is displayed:
Input comment for this setup memory. (Max 20 char)
Type desired comment (maximum 20 characters), then press the
press
key without typing a comment, comment line is cleared.
Enter
To change the E5250A setup to specified setup memory
Select the GOTO SETUP softkey. The following message is displayed:
Input setup memory number to be connected?
Type the desired setup memory number, then press the
Memory changes to the specified setup memory, and the E5250A changes to the new
settings.
To cancel this operation, press the
To change the settings in a setup memory
Select the CHANGE CHAN softkey. A cursor for selecting input-output crosspoints
is displayed. And the following softkeys are displayed for changing the settings of
the setup memory.
Change the setting of the setup memory by using these softkeys.
Invalid Input Port
The following E5250A input port combinations cannot be used for making a setup
memory.
IV5 (INPUT 5) and HF1 (INPUT 7) and CV1 (INPUT 9)
IV6 (INPUT 6) and HF2 (INPUT 8) and CV2 (INPUT 10)
For example, when IV5 is used, HF1 and CV1 cannot be used.
Agilent E5250A User's Guide, Edition 11
Using Virtual Front Panel Utility
Enter
key without typing a number.
Enter
Controlling the E5250A
key. If you
Enter
key. Current Setup
5-39

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Agilent e5252aAgilent e5255a